Stimulator control bits and relay switching bits not working after power up
Tech Note: TN0220
Product: RX7, MS4, MS16
Version: TBD
Date Added: 2006-09-22
Issue
After power up, the stimulus isolator control bits and relay switching control bits do not work.
Workaround
To restore full functionality, execute a hardware reset on the RX7 using zBusMon.